Display: 24000 counts
|
? |
Measure loop current signals with very high accuracy and
0.001 mA resolution
|
? |
Measure voltage output process signals from PLCs, transmitters
|
? |
Source or simulate 24 mA loop current
|
? |
Source voltage to 10V
|
? |
24V loop supply with simultaneous current measurement |
Function |
Range |
Resolution |
DC mA Input
|
0 ~ 24 mA |
0.001 mA |
DC mA Output
|
0 ~ 24 mA |
0.001 mA |
Specification
Specification are based on a one year calibration cycle and apply from +18℃ to +28℃ unless stated otherwise. ?Counts? means number of increments or decrements of the least significant digit.
DC V Input and Output
Range |
Resolution |
Accuracy ±(% of reading + Counts) |
100 mV |
0.01 mV |
0.02 % + 2 |
10 V
|
0.001 V |
0.02 % + 2 |
Input impedance: 2MΩ
(nominal),< 100pF
Over voltage protection: 30 V
Voltage driver capability: 1 mA |
DC mA Input and Output
Range |
Resolution |
Accuracy ±(% of reading + Counts) |
24 mA |
0.001mA |
0.015 % +4 |
Overload protection:
125 mA,250V
fast acting fuse
Percent display:
0%=4mA,100%=20mA
Source mode:
compliance 1000Ω at 20mA for battery voltage ≥6.8V,
(700Ω at 20mA for battery voltage 5.8 to 6.8V)
Simulate mode: External loop voltage requirement: 24V nominal, 30V maximum, 12V minimum. |
LOOP POWER
General Specifications:
Maximum voltage applied between any jack and earth ground or between any tow jack:30V
Storage temperature:-40℃~60℃
Operating temperature:-10℃~55℃
Operating altitude:3000 meters maximum
Temperature coefficient:±0.005% of range per ℃ for the temperature range -10℃ to 18℃ and 28℃ to 55℃
Relative humidity:95% up to 30℃,75% up to 40℃,45% up to 50℃,35% up to 55℃
Shock:Random 2g , 5Hz to 500Hz
Safety:1 meter drop test
Power requirements:9VDC, 006P or 1604A
Size:190mm L × 89mm W × 42mm H
Weight: 350g (include battery)